Spectrum Analyzer Commands :CALCulate Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-17
Limit Line Y Value
:CALCulate:LIMit:Y
Description: Sets the location of the active limit line on the y-axis at the specified
location. This moves the entire limit and moves the current active limit
point by the given value. <y-parameter> is defined in the current
y-axis. Note that this changes the Move Limit on the front panel to
Limit if it is currently set to Point. Use :CALCulate:LIMit:TYPe to
set the currently active limit line.
Syntax: :CALCulate:LIMit:Y <y-parameter>
Cmd Parameter: <y-parameter>
Query Response: <y-parameter>
Default Unit: Current y-axis unit.
Example: To move limit and set the active point to 5 dbm:
:CALCulate:LIMit:Y 5
(If y-axis unit is dBm)
:CALCulate:LIMit:Y 5dBm
Front Panel
Access: Shift-6 (Limit), Limit Edit, Amplitude
Set Default Limit
:CALCulate:LIMit[:SET]:DEFault
Description: Deletes all limit points for the currently active limit line and sets the
default limit line value. Note that this command turns on the currently
selected limit line if it is not already on. The current selected limit line
can be modified by using the command :CALCulate:LIMit:TYPe.
Syntax: :CALCulate:LIMit[:SET]:DEFault
Front Panel
Access: Shift-6 (Limit), Set Default Limit