:CALCulate:LIMit Subsystem VNA Commands
3-30 PN: 10580-00306 Rev. E MS20xxC PM
Lower Limit State
:CALCulate<Tr>:LIMit:LOWer[:STATe]
Description: Turns the lower limit line of the given trace <Tr> ON or OFF. <Tr> is
the trace number in the range 1 to 4. If no trace number is specified,
then the <Tr> parameter defaults to trace number 1. <Tr> is the trace
number in the range 1 to 4. If no trace number is specified, then the
<Tr> parameter defaults to trace number 1. The query version of the
command returns a 1 if the lower limit line of the given trace is ON, and
returns a 0 if it is OFF.
Syntax: :CALCulate<Tr>:LIMit:LOWer[:STATe] OFF|ON|0|1
:CALCulate<Tr>:LIMit:LOWer[:STATe]?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF or 0 (query returns 0 for OFF)
Example: To turn on lower limit of trace 1
:CALCulate:LIMit:LOWer ON
:CALCulate1:LIMit:LOWer 1
:CALCulate:LIMit:LOWer:STATe ON
To turn off upper limit of trace 4
:CALCulate4:LIMit:LOWer OFF
:CALCulate4:LIMit:LOWer 0
:CALC4:LIM:LOW:STAT 0
Related Command: :CALCulate:LIMit[:STATe]
Front Panel
Access: Shift 6 (Limit), Limit State