VNA Commands :Display Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-83
Scale Resolution Per Division
:DISPlay:WINDow:TRACe<Tr>:Y[:SCALe]:PDIVision
Description: Sets the scale per division for the y-axis. For Group Delay, sets the scale
(time/division) for the y-axis. For Phase, sets the scale (degree/division)
for the y-axis. For Log Magnitude, Log Mag/2, and Log Polar, sets the
scale (dB/division) for the y-axis. For Real Impedance and Imaginary
Impedance, sets the scale (ohm/division) for the y-axis. For all other
measurements, the y-axis is unitless. <Tr> is the trace number in the
range 1 to 4. If no trace number is specified, then the <Tr> parameter
defaults to trace number 1.
Syntax: :DISPlay:WINDow:TRACe<Tr>:Y[:SCALe]:PDIVision <value>
:DISPlay:WINDow:TRACe<Tr>:Y[:SCALe]:PDIVision?
Cmd Parameter: <NRf> <value> (depends on display type)
Query Response: <NR3> <value> (depends on display type)
Range: Log Magnitude: 0.05 dB to 40 dB
Phase: 0.1° to 90°
SWR: 0.001 to 10
Group Delay: 1 ps to 260 ns
Real: 0.01 to 260
Imag: 0.01 to 260
Log Mag/2: 0.05 dB to 40 dB
Smith Chart: 1 to 260
Inverted Smith Chart: 1 to 260
Log Polar: 0.05 dB to 40 dB
Linear Polar: 0.001 to 26
Real Impedance: 0.01 ohm to 100000 ohm
Imaginary Impedance: 0.01 ohm to 100000 ohm
Default Value: Log Magnitude: 10 dB
Phase: 45°
SWR: 1
Group Delay: 1 ns
Real: 0.2
Imag: 0.2
Log Mag/2: 10 dB
Smith Chart: 10
Inverted Smith Chart: 10
Log Polar: 10 dB
Linear Polar: 0.2
Real Impedance: 10 ohm
Imaginary Impedance: 10 ohm