EasyManua.ls Logo

Anritsu MS2026C - Trace a Operation

Anritsu MS2026C
410 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
:TRACe Subsystem Spectrum Analyzer Commands
6-62 PN: 10580-00306 Rev. E MS20xxC PM
Trace A Operation
:TRACe1:OPERation
Description: This command specifies how successive traces are combined to produce
the resulting display values.
Setting the operation to NORMal is equivalent to pressing
Shift-5 (Trace), Trace A Operations, NormalA on the front panel. This
displays a trace based on the detection method selected.
Setting the operation to MAXHold is equivalent to pressing
Shift-5 (Trace), Trace A Operations, Max HoldA on the front panel.
This displays the largest signal for each display point over multiple
sweeps.
Setting the operation to MINHold is equivalent to pressing
Shift-5 (Trace), Trace A Operations, Min HoldA on the front panel. This
displays the smallest signal for each display point over multiple sweeps.
Setting the operation to AVERage is equivalent to pressing
Shift-5 (Trace), Trace A Operations, AverageA on the front panel. This
displays the average value of multiple sweeps for each display point.
The query version of the command returns the current operation mode
or “NONE” if no operation is set.
Syntax: :TRACe1:OPERation NORMal|MAXHold|MINHold|AVERage
:TRACe1:OPERation?
Cmd Parameter: <char> NORMal|MAXHold|MINHold|AVERage
Query Response: <char> NORM|MAXH|MINH|AVER
Default Value: NORMal
Related Command: [:SENSe]:AVERage:TYPE
Front Panel
Access: Shift-5 (Trace), Trace A Operations, NormalA
Shift-5 (Trace), Trace A Operations, Max HoldA
Shift-5 (Trace), Trace A Operations, Min HoldA
Shift-5 (Trace),
Trace A Operations, AverageA

Table of Contents