:CALCulate<Tr>:FILTer[:GATE]:TIME Subsystem VNA Commands
3-14 PN: 10580-00306 Rev. E MS20xxC PM
Time Domain Gate Notch State
:CALCulate<Tr>:FILTer[:GATE]:TIME:NOTCh
Description: Sets the gate into notch mode for the given trace. Setting the value
to ON or to 1 turns on the gate into a notch (in other words, the gate
suppresses rather than passes through the time domain data within the
start/stop gate settings). Setting the value to OFF or to 0 turns off the
gate notch mode. The query version of this command returns 1 if gate
notch is on, otherwise returns 0 for off. Note that if gate coupling is on,
then setting the gate notch of one trace also sets the gate notch of all
other traces. Note that setting this also sets the given trace as the active
trace if it is not already active.
Note that the command :CALC<Tr>:FILT[:GATE]:DIST:NOTC
functions the same as this command.
Syntax: :CALCulate<Tr>:FILTer[:GATE]:TIME:NOTCh ON|OFF|1|0
:CALCulate<Tr>:FILTer[:GATE]:TIME:NOTCh?
Cmd Parameter: <boolean> ON|OFF|1|0
Query Response: <bNR1> 1|0
Default Value: 0
Example: To set the gate notch for Trace 3 to OFF:
:CALC3:FILT:GATE:TIME:NOTC 0
OR
:CALC3:FILT:GATE:TIME:NOTC OFF
Related Command: :CALC<Tr>:FILT[:GATE]:DIST:NOTC
Front Panel
Access: Freq/Time/Dist, Gate, Gate Notch