:TRIGger Subsystem Spectrum Analyzer Commands
6-66 PN: 10580-00306 Rev. E MS20xxC PM
Video Trigger Position (time)
:TRIGger[:SEQuence]:VIDeo:DELay
Description: This command sets the video triggering delay as either a percentage of
the display or in time units. If setting the delay by time is desired, then
time units must be specified when sending the command. The query
version of this command returns the video triggering delay as a
percentage
Syntax: :TRIGger[:SEQuence]:VIDeo:DELay <percentage> or <time>
:TRIGger[:SEQuence]:VIDeo:DELay?
Cmd Parameter: <percentage> or <time>
Query Response: <percentage> or <time>
Range: 100% to +200% (–1 ms to +2 ms)
Default Value: –1
Default Unit: %
Example: To set the delay to 1 ms:
:TRIGger:SEQuence:VIDeo:DELay 1 ms
To set the delay to 1%:
:TRIGger:SEQuence:VIDeo:DELay 1
Front Panel
Access: Shift-3 (Sweep), Triggering, Source, Free Run|External|Video
Video Trigger Level
:TRIGger[:SEQuence]:VIDeo:LEVel
Description: This command sets the video triggering level.
Syntax: :TRIGger[:SEQuence]:VIDeo:LEVel <amplitude>
:TRIGger[:SEQuence]:VIDeo:LEVel?
Cmd Parameter: <amplitude>
Query Response: <amplitude>
Range: +30 dBm to –150 dBm
Default Value: 65.0 dBm
Default Unit: Current amplitude unit
Front Panel
Access: Shift-3 (Sweep), Triggering, Level