VNA Commands [:SENSe]:CORRection:COLLect Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-129
Configure Calibration Type
[:SENSe]:CORRection:COLLect:CTYPe
Description: Configures the calibration types. <cal type1> specifies the type of
calibration (1-port, 2-port, response, etc) and must be one of the
following values:
RF2P|RFP1|RFP2|RFBP|TRFP|TRRP|TRBP|RRP1|RRP2|RRBP|
|2PFP|2PRP
Refer to the table of calibration types at
“[:SENSe]:CORRection:COLLect:TYPE” on page 3-130. Command
[:SENSe]:CORRection:COLLect:TYPe has been kept for backward
compatibility.
This command, [:SENSe]:CORRection:COLLect:CTYPe, allows you
to specify whether the cal is Standard or Flex in addition to the
Cal type.
<cal type2> specifies if the instrument performs a standard or flex
type calibration, and must use one of the following values:
FLEX|STANdard
Syntax: [:SENSe]:CORRection:COLLect:CTYPe <cal type>
FLEX|STANdard
[:SENSe]:CORRection:COLLect:CTYPE?
Cmd Parameter: <char>
RF2P|RFP1|RFP2|RFBP|TRFP|TRRP|TRBP|RRP1|RRP2|RRBP|2PFP|
2PRP, FLEX|STANdard
Query Response: <char>
RF2P|RFP1|RFP2|RFBP|TRFP|TRRP|TRBP|RRP1|RRP2|RRBP|2PFP|
2PRP, FLEX|STAN
Default Value: RF2P, STAN
Example: To set the cal type to Full S
11
, Flex:
:SENS:CORR:COLL:CTYP RFP1, FLEX
To query for the current cal type:
:SENS:CORR:COLL:CTYP?
The response would be: RFP1, FLEX
Front Panel
Access: Shift-2 (Calibrate), Cal Type