:SOURce:CORRection:RVELocity Subsystem VNA Commands
3-150 PN: 10580-00306 Rev. E MS20xxC PM
Cutoff Freq
:SOURce:CORRection:RVELocity:FCUToff
Description: Sets the Cutoff Frequency for DTF measurements.
Syntax: :SOURce:CORRection:RVELocity:FCUToff <freq>
:SOURce:CORRection:RVELocity:FCUToff?
Parameter: <freq>
Cmd Parameter: <NRf> <number> (hertz)
Query Response: <NR3> <number> (hertz)
Range: 5 kHz to 20 GHz for MS2028C, MS2038C
5 kHz to 15 GHz for MS2027C, MS2037C
5 kHz to 6 GHz for MS2026C, MS2036C
Default Value: 5000 Hz
Default Unit: Hz
Front Panel
Access: Freq/Time/Dist (or Freq), Domain Setup, Setup Distance,
Additional Dist Setup, Cutoff Freq (when DUT Line Type is Waveguide)
DUT Line Type
:SOURce:CORRection:RVELocity:MEDium
Description: Sets the calibration line type.
Syntax: :SOURce:CORRection:RVELocity:MEDium COAX|WGUide
:SOURce:CORRection:RVELocity:MEDium?
Cmd Parameter: <char> [COAX|WGUide]
Query Response: <char> [COAX|WGU]
Default Value: COAX
Front Panel
Access: Freq/Time/Dist (or Freq), Domain Setup, Setup Distance, Additional
Dist Setup, DUT Line Type