EasyManua.ls Logo

Anritsu MS2026C - READ Subsystem; Read Adjacent Channel Power Ratio

Anritsu MS2026C
410 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
:READ Subsystem Spectrum Analyzer Commands
6-50 PN: 10580-00306 Rev. E MS20xxC PM
6-11 :READ Subsystem
This set of commands combines the :ABORt, :INITiate and :FETCh commands. It
aborts any current triggering sequence and sets the trigger state to idle. It then initiates a
new active measurement (in other words, begins the collection of new data). When the
measurement is complete, it returns the result. These commands do not switch to another
measurement.
To get the current measurement data, use the :FETCh command.
Read Adjacent Channel Power Ratio
:READ:ACPower?
Description: Triggers a new adjacent channel power ratio measurement and returns
the results: main channel power, lower adjacent channel power, and
upper adjacent channel power. It is a combination of the commands
:ABORT; :INITiate; :FETCh:ACPower? The channel power
measurement must be the active measurement (specified by the
command :CONFigure:ACPower). The current measurement can be
queried using the command :CONFigure?
Data is returned as 3 comma-separated values: main channel power,
lower adjacent channel power, upper adjacent channel power.
Syntax: :READ:ACPower?
Cmd Parameter: NA (query only)
Related Command: :READ:ACPower?
:CONFigure