:CALCulate:LIMit Subsystem VNA Commands
3-42 PN: 10580-00306 Rev. E MS20xxC PM
Upper Limit Next Point Left
:CALCulate<Tr>:LIMit:UPPer:POINt:LEFT
Description: Sets the limit point to the left of the upper limit active point of the given
trace <Tr> as the new active point. <Tr> is the trace number in the
range 1 to 4. If no trace number is specified, then the <Tr> parameter
defaults to trace number 1.
Syntax: :CALCulate<Tr>:LIMit:UPPer:POINt:LEFT
Cmd Parameter: NA
Query Response: NA (no query)
Example: To make the upper limit point to the left of the current active point of
trace 2 as the new active point:
:CALCulate2:LIMit:UPPer:POINt:LEFT
:CALC2:LIM:UPP:POIN:LEFT
Related Command: :CALCulate<Tr>:LIMit:UPPer:POINt:RIGHt
Front Panel
Access: Shift 6 (Limit)), Limit Edit, Next Point Left
Upper Limit Next Point Right
:CALCulate<Tr>:LIMit:UPPer:POINt:RIGHt
Description: Sets the limit point to the right of the upper limit active point of the
given trace <Tr> as the new active point. <Tr> is the trace number in
the range 1 to 4. If no trace number is specified, then default trace is
trace number 1.
Syntax: :CALCulate<Tr>:LIMit:UPPer:POINt:RIGHt
Cmd Parameter: NA
Query Response: NA (no query)
Example: To make the upper limit point to the right of the current active point of
trace 2 as the new active point:
:CALCulate2:LIMit:UPPer:POINt:RIGHt
:CALC2:LIM:UPP:POIN:RIGH
Related Command: :CALCulate<Tr>:LIMit:UPPer:POINt:LEFT
Front Panel
Access: Shift 6 (Limit), Limit Edit, Next Point Right