[:SENSe]:CORRection:COLLect Subsystem VNA Commands
3-134 PN: 10580-00306 Rev. E MS20xxC PM
Calibration Status
[:SENSe]:CORRection:COLLect:STATus?
Description: This command requests information about the calibration status. The
command returns 0 if none, 1 if calibration has already started, 2 if
calibration has been aborted, 3 if a calibration is currently calculating,
and 4 if a calibration has been completed.
Syntax: [:SENSe]:CORRection:COLLect:STATus?
Cmd Parameter: NA (query only)
Query Response: <NR1> (integer)
0 = none
1 = started
2 = aborted
3 = calculating
4 = completed
Front Panel
Access: NA
Calibration Accuracy Status
[:SENSe]:CORRection:COLLect:STATus:ACCuracy?
Description: This command requests information about the calibration accuracy
status. The command returns 0 when no calibration is available
(CAL off), 1 when the calibration accuracy is high
(OK: Accuracy High), 2 when the accuracy is moderate due to a
change in power level (?P: Accuracy Moderate), 3 when the accuracy is
moderate due to a change in temperature level by more than 5 deg C
(?T: Accuracy Moderate), and 4 when the accuracy is low due to a
change in temperature level by more than 10 deg C (X: Accuracy Low).
Syntax: [:SENSe]:CORRection:COLLect:STATus:ACCuracy?
Cmd Parameter: NA (query only)
Query Response: <NR1> (integer)
0 = CAL off
1 = OK: Accuracy High
2 = ?P: Accuracy Moderate
3 = ?T: Accuracy Moderate
4 = X: Accuracy Low
Front Panel
Access: Shift2 (Calibrate), Existing Cal Info