:READ Subsystem Spectrum Analyzer Commands
6-52 PN: 10580-00306 Rev. E MS20xxC PM
Read Occupied Bandwidth
:READ:OBWidth?
Description: Triggers a new occupied bandwidth measurement and returns the
results: occupied bandwidth, percent of power and dB down. It is a
combination of the commands :ABORT; :INITiate;
:FETCh:OBWidth? The occupied bandwidth measurement must be the
active measurement (specified by :CONFigure:OBWidth). The current
measurement can be queried using :CONFigure?
Data is returned as 3 comma-separated values: occupied bandwidth,
percent of power, dB down.
Syntax: :READ:OBWidth?
Cmd Parameter: NA (query only)
Default Unit: For OBW: Hz
For Percent of Power: %
For dB Down: dB
Related Command: :CONFigure