:MMEMory Subsystem All Mode Commands
8-6 PN: 10580-00306 Rev. E MS20xxC PM
Transfer Data
:MMEMory:DATA?
Description: Transfers the data stored in the given file from the instrument to the
controlling program. Data is transferred in the form of
<header><block>. The ASCII <header> specifies the number of
data bytes. It appears as #AX, where A is the number of digits in X,
and X is the number of bytes in <block>.
<file name> should be enclosed in either single quotes (‘ ’) or double
quotes (“ ”) and should contain a file extension (.jpg, for example). The
file must not be larger than 524288 bytes. Use the command
:MMEMory:MSIS to set the current storage location.
Syntax: :MMEMory:DATA? <file name>
Cmd Parameter: NA (query only)
Query Response: <string> <file name>
Related Command: :MMEMory:MSIS INTernal|USB
Front Panel
Access: NA
Delete Data/Location
:MMEMory:DELete
Description: This command deletes a file. <file name> must be enclosed in either
single quotes (‘ ‘) or double quotes (“ “) and must include the file
extension.
Syntax: :MMEMory:DELete <file name>
Cmd Parameter: <string> <file name>
Query Response: NA (no query)
Front Panel
Access: Shift-7 (File), Delete, Select or De-Select, Delete