Spectrum Analyzer Commands :CALCulate Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-7
Lower Limit Alarm
:CALCulate:LIMit:LOWer:ALARm
Description: Enables/disables the lower limit alarm. It is a combination of the
commands :CALCulate:LIMit:TYPe 1;
:CALCulate:LIMit:ALARm ON|OFF. Setting the value to ON or 1
turns on the lower limit alarm. Setting the value to OFF or 0 turns off
the lower limit alarm. The query version of the command returns a 1 if
the lower limit line alarm is ON and returns 0 if OFF. Note that using
this command sets the lower limit line to be active for editing.
Syntax: :CALCulate:LIMit:LOWer:ALARm OFF|ON|0|1
:CALCulate:LIMit:LOWer:ALARm?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF
Related Command: :CALCulate:LIMit:ALARm
Front Panel
Access: Shift-6 (Limit), Limit Alarm
Lower Limit Fail State
:CALCulate:LIMit:LOWer:FAIL?
Description: Query whether the lower limit line has failed or not. The command
returns a 0 on success, returns a 1 on fail, and returns a 2 if the lower
limit line is OFF or if the alarm is OFF.
Syntax: :CALCulate:LIMit:LOWer:FAIL?
Front Panel
Access: NA
Number of Lower Limit Points
:CALCulate:LIMit:LOWer:POINt?
Description: Returns the number of points currently in the lower limit line.
Syntax: :CALCulate:LIMit:LOWer:POINt?
Default Value: 2
Related Command: :CALCulate:LIMit:POINt?
Front Panel
Access: NA