EasyManua.ls Logo

Anritsu MS2026C - 3-31 [:Sense]:Trace Subsystem; Trace Domain

Anritsu MS2026C
410 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VNA Commands [:SENSe]:TRACe Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-145
3-31 [:SENSe]:TRACe Subsystem
This subsystem includes commands that provide general settings for each trace.
Trace Domain
[:SENSe]:TRACe<Tr>:DOMain
Description: Defines the domain for the given trace <Tr>. <Tr> is the trace
number in the range 1 to 4. If no trace number is specified, then the
<Tr> parameter defaults to trace number 1. The query version of this
command returns “FREQ” if domain is Frequency, “TIME” if domain
is Time, “DIST” if domain is distance, and “FGT” if domain is
Frequency Gated in Time.
Syntax: [:SENSe]:TRACe<Tr>:DOMain FREQuency|TIME|DISTance|FGT
[:SENSe]:TRACe<Tr>:DOMain?
Cmd Parameter: <char> FREQuency|TIME|DISTance|FGT
Query Response: <char> FREQ|TIME|DIST|FGT
Default Value: Trace 1: FREQ
Trace 2: FREQ
Trace 3: FREQ
Trace 4: FREQ
Example: To assign Time domain to trace 2:
:SENSe:TRACe2:DOMain TIME
Front Panel
Access: Measure, Domain Selection

Table of Contents