:CALCulate:MARKer Subsystem VNA Commands
3-56 PN: 10580-00306 Rev. E MS20xxC PM
Marker Readout Format
:CALCulate:MARKer:DISPlay:FORMat
Description: Sets the display readout format for markers. The query version of the
command returns “NONE” if the display readout format is set to None,
“SCRE” if Screen, “TABL” if Table, and “TRAC” if Trace.
Syntax: :CALCulate:MARKer:DISPlay:FORMat
NONE|SCREen|TABLe|TRACe
:CALCulate:MARKer:DISPlay:FORMat?
Cmd Parameter: <char> NONE|SCREen|TABLe|TRACe
Query Response: <char> NONE|SCRE|TABL|TRAC
Default Value: NONE
Example: To set marker readout format to Table:
:CALCulate:MARKer:DISPlay:FORMat TABLe
:CALCulate:MARKer:DISPlay:FORMat TABL
:CALC:MARK:DISP:FORM TABL
Front Panel
Access: Marker, Readout Format
Marker Domain Type
:CALCulate:MARKer<Mk>:DOMain?
Description: Query the specified marker <Mk> domain type. <Mk> is the marker
number in the range of 1 to 8. If no marker number is specified, then the
marker number (the <Mk> value) defaults to 1. This command returns
“FREQ” if the specified marker domain is frequency, “TIME” if time,
and “DIST” if distance.
Syntax: :CALCulate:MARKer<Mk>:DOMain?
Cmd Parameter: NA (query only)
Query Response: <char> FREQ|TIME|DIST
Front Panel
Access: NA