VNA Commands :CALCulate:LIMit Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-25
Add Lower Limit Point
:CALCulate<Tr>:LIMit:LOWer:POINt:ADD
Description: Adds a new limit point to the lower limit line of the given trace <Tr>.
<Tr> is the trace number in the range 1 to 4. If no trace number is
specified, then the <Tr> parameter defaults to trace number 1.
Syntax: :CALCulate<Tr>:LIMit:LOWer:POINt:ADD
Cmd Parameter: NA
Query Response: NA (no query)
Example: To add a point to the lower limit line on trace 2:
:CALC2:LIM:LOW:POIN:ADD
Front Panel
Access: Shift 6 (Limit), Limit Edit, Add Point
Delete Lower Limit Point
:CALCulate<Tr>:LIMit:LOWer:POINt:DELete
Description: Deletes the lower limit point of the given trace <Tr>. After deletion,
the point that is immediately to the left of the point that was deleted
becomes the active point. <Tr> is the trace number in the range 1 to 4.
If no trace number is specified, then the <Tr> parameter defaults to
trace number 1. Note that deletion is valid only if more than 2 limit
points are present.
Syntax: :CALCulate<Tr>:LIMit:LOWer:POINt:DELete
Cmd Parameter: NA
Query Response: NA (no query)
Example: To delete the trace 4 lower limit current active point:
:CALCulate4:LIMit:LOWer:POINt:DELete
Front Panel
Access: Shift 6 (Limit), Limit Edit, Delete Point