:CALCulate:TRANsform Subsystem VNA Commands
3-78 PN: 10580-00306 Rev. E MS20xxC PM
Band Pass Mode Response
:CALCulate<Tr>:TRANsform:TIME:BPASs:STIMulus
Description: Sets the response type to be used in the band pass transformation
process for the given trace. The response type is set to either Standard
or Phasor Impulse. <Tr> is the trace number in the range 1 to 4. If no
trace number is specified, then the <Tr> parameter defaults to trace
number 1. The query version of this command returns the string STAN
for Standard (no phasor impulse) and PHAS for phasor impulse.
Syntax: :CALCulate<Tr>:TRANsform:TIME:BPASs:STIMulus
STANdard|PHASor
:CALCulate<Tr>:TRANsform:TIME:BPASs:STIMulus?
Cmd Parameter: <char> STANdard|PHASor
Query Response: <char> STAN|PHAS
Default Unit: STAN
Example: To Set Trace 1 to Phasor Impulse:
:CALCulate:TRANsform:TIME:BPASs:STIMulus PHASor
OR:
:CALC1:TRAN:TIME:BPAS:STIM PHAS
Related Command: :CALCulate:TRANsform:TIME:TYPE:AUTO
:CALCulate:TRANsform:TIME:TYPE?
Front Panel
Access: Shift-4 (Measure), Domain Selection, Band Pass Response
(Note that access via this key sequence requires that the domain
processing be set to “Band Pass”.)