:MMEMory Subsystem Spectrum Analyzer Commands
6-48 PN: 10580-00306 Rev. E MS20xxC PM
Save Limit
:MMEMory:STORe:LIMit
Description: Stores the current limit setup into the file specified by <file name>.
<file name> should be enclosed in either single quotes (‘ ’) or double
quotes (“ ”) and should not contain a file extension. Use the command
:MMEMory:MSIS to set the current storage location.
Syntax: :MMEMory:STORe:LIMit <file name>
Cmd Parameter: <file name>
Related Command: :MMEMory:MSIS INTernal|USB
Front Panel
Access: Shift-7 (File), Save, Change Type, (select Limit Line from list)
Save Setup
:MMEMory:STORe:STATe
Description: Stores the current setup into the file that is specified by
<file name>. <file name> should be enclosed in either single
quotes (‘ ’) or double quotes (“ ”) and should not contain a file extension.
Use the command :MMEMory:MSIS to set the current storage location.
The <integer> parameter is not currently used, but it must be sent.
Send a value of 0.
Syntax: :MMEMory:STORe:STATe <integer>,<file name>
Cmd Parameter: <integer>, <file name>
Related Command: :MMEMory:LOAD:STATe
:MMEMory:MSIS INTernal|USB
Front Panel
Access: Shift-7 (File), Save, Change Type (select Setup from list)