:CALCulate Subsystem Spectrum Analyzer Commands
6-16 PN: 10580-00306 Rev. E MS20xxC PM
Limit X Value
:CALCulate:LIMit:X
Description: Sets the location of the active limit line on the x-axis at the specified
location. This moves the entire limit and moves the active limit point to
the given value. <x-parameter> is defined in the current x-axis. Note
that this changes the Move Limit on the front panel to Limit if it is
currently set to Point. Use :CALCulate:LIMit:TYPe to set the
currently active limit line.
Syntax: :CALCulate:LIMit:X <x-parameter>
Cmd Parameter: <x-parameter>
Query Response: <x-parameter>
Default Unit: Hz or for zero span in seconds
Example: To move the limit and set active point to 5 Hz:
:CALCulate:LIMit:X 5
:CALCulate:LIMit:X 5Hz
To move the limit and set active point to 500 MHz:
:CALCulate:LIMit:X 500MHz
To move the limit and set active point to 2.5 seconds (In zero span):
:CALCulate:LIMit:X 2.5
:CALCulate:LIMit:X 2.5s
To move the limit and set active point to 25 microseconds (In zero span):
:CALCulate:LIMit:X 25us
Front Panel
Access: Shift-6 (Limit), Limit Edit, Frequency