Spectrum Analyzer Commands [:SENSe] Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-77
Input Impedance
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]
Description: Sets the input impedance that is used for amplitude correction and
conversion between units (dBm versus dBV versus Volts, and so forth).
If the value of <integer> is 50, then no correction is performed. If the
value of <integer> is 75, then correction is based on Anritsu adapter
12N50-75B.
To place the instrument in Other or offset mode, send a 2. After the
instrument is in the “Other” mode, the command,
[:SENSe]:CORRection:IMPedance[:INPut]:OFFSet, can be used
to adjust the offset.
Syntax: [:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]
<integer>
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]?
Cmd Parameter: <integer>
Query Response: <integer>
Range: 50 ohm or 75 ohm, all other values are treated as described in command
description:
Default Value: 50 ohm
Front Panel
Access: Shift-8 (System), Application Options, Impedance