:MMEMory:STORe Subsystem VNA Commands
3-100 PN: 10580-00306 Rev. E MS20xxC PM
Save Measurement
:MMEMory:STORe:TRACe
Description: Stores the trace into the file that is specified by <file name>.
<file name> must be enclosed in either single quotes (‘ ’) or double
quotes (“ ”) and must not contain a file extension. The <integer>
parameter is used to distinguish which type of files to save. The
following types are available:
Syntax: :MMEMory:STORe:TRACe <integer>,<file name>
Cmd Parameter: <integer>,<string> (1|2, filename)
Query Response: NA (no query)
Example: To save the trace into the file named “trace”.
:MMEMory:STORe:TRACe 1,”trace”
Related Command: :MMEMory:LOAD:TRACe
Front Panel
Access: Shift-7 (File), Save, Change Type (select file type from list)
Shift-7 (File), Save Measurement
Note
The integer parameters that are used in this command are specific to the vector
network analyzer modes, not for spectrum analyzer mode.
<Integer> : File type
1 : Measurement file (default, if number is not 2 to 6)
2 : S2P Real/Imag
3 : S2P Lin Mag/Phase
4 : S2P Log Mag/Phase
5:Text
6:CSV