:CALCulate Subsystem Spectrum Analyzer Commands
6-12 PN: 10580-00306 Rev. E MS20xxC PM
Limit Point Y Value
:CALCulate:LIMit:POINt:Y
Description: Sets the location of the active limit point on the y-axis at the specified
location. <y-parameter> is defined in the current y-axis. Note that
this changes the Move Limit on the front panel to Point if it is currently
set to Limit. The query version of the command returns the location of
the active limit point on the y-axis. Use :CALCulate:LIMit:TYPe to
set the currently active limit line.
Syntax: :CALCulate:LIMit:POINt:Y <y-parameter>
:CALCulate:LIMit:POINt:Y?
Cmd Parameter: <y-parameter>
Query Response: <y-parameter>
Default Unit: Current y-axis unit.
Example: To set the active point to 5 dBm:
:CALCulate:LIMit:POINt:Y 5
(If y-axis unit is dBm)
:CALCulate:LIMit:POINt:Y 5dBm
Front Panel
Access: Shift-6 (Limit), Limit Edit, Amplitude
Number of Limit Points
:CALCulate:LIMit:POINt?
Description: Returns the number of points currently in the selected limit line. Use
:CALCulate:LIMit:TYPe to set the currently active limit line.
Syntax: :CALCulate:LIMit:POINt?
Related Command: :CALCulate:LIMit:TYPe
Front Panel
Access: NA