Spectrum Analyzer Commands [:SENSe] Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-71
Trace Mode (Normal/Average/Max Hold/Min Hold)
[:SENSe]:AVERage:TYPE
Description: Specifies how successive traces are combined to produce the resulting
display value. Setting the TYPE to NONE is the equivalent of setting
the trace mode to “NormalA” on the front panel. The displayed value
for a point is the current measured value for that point. Setting the
TYPE to SCALar is the equivalent of setting the trace mode to
“AverageA” on the front panel. The displayed value for a point is the
average of the last <integer> measured values where <integer> is set
by [:SENSe]:AVERage:COUNt. Setting the TYPE to MAXimum is the
equivalent of setting the trace mode to “Max HoldA” on the front
panel. The displayed value for a point is the maximum measured value
for that point over sweeps. Setting the TYPE to MINimum is the
equivalent of setting the trace mode to “Min HoldA” on the front panel.
The displayed value for a point is the minimum measured value for that
point over sweeps.
Syntax: [:SENSe]:AVERage:TYPE NONE|SCALar|MAXimum|MINimum
[:SENSe]:AVERage:TYPE?
Cmd Parameter: <char> NONE|SCALar|MAXimum|MINimum
Query Response: <char> NONE|SCAL|MAX|MIN
Default Value: NONE
Example: To set the TYPE to SCALar:
:SENSe:AVERage:TYPE SCALar
To set the TYPE to MAXimum:
:SENSe:AVERage:TYPE MAXimum
Related Command: [:SENSe]:AVERage:COUNt
Front Panel
Access: Shift-5 (Trace), Trace A Operations