VNA Commands :CALCulate:TRANsform Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-69
Start Distance
:CALCulate:TRANsform:DISTance:STARt
Description: Sets the start distance for DTF measurements. The query version of
this command returns the start distance in millimeters if the current
distance unit is set to meter, and otherwise returns the start distance in
feet.
Syntax: :CALCulate:TRANsform:DISTance:STARt
:CALCulate:TRANsform:DISTance:STARt?
Cmd Parameter: <NRf> (meters or feet)
Query Response: <NR3> (millimeters or feet)
Range: –3000.0 m to +3000.0 m
Default Value: 0.0 mm
Default Unit: Meters (m) when setting, Millimeters (mm) for query
Example: To set the start distance to 5 meters:
:CALC:TRAN:DIST:STAR 5
To set the start distance to 6 millimeters:
:CALCulate:TRANsform:DISTance:STARt 6mm
Front Panel
Access: Freq/Time/Dist, Start Dist
Stop Distance
:CALCulate:TRANsform:DISTance:STOP
Description: Sets the stop distance for DTF measurements. The query version of this
command returns the stop distance in millimeters if the current
distance unit is set to meter, and otherwise returns the stop distance in
feet.
Syntax: :CALCulate:TRANsform:DISTance:STOP
:CALCulate:TRANsform:DISTance:STOP?
Cmd Parameter: <NRf> (meters or feet)
Query Response: <NR3> (millimeters or feet)
Range: –3000.0 m to +3000.0 m
Default Value: 6850 mm
Default Unit: Meters (m) when setting, Millimeters (mm) for query
Front Panel
Access: Freq/Time/Dist, Stop Dist