[:SENSe] Subsystem Spectrum Analyzer Commands
6-76 PN: 10580-00306 Rev. E MS20xxC PM
Channel Power Measurement State
[:SENSe]:CHPower:STATe
Description: Sets the state of the channel power measurement, ON or OFF. When
using :CONFigure:CHPower, the state is automatically set to ON.
Syntax: [:SENSe]:CHPower:STATe OFF|ON|0|1
[:SENSe]:CHPower:STATe?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF
Related Command: :CONFigure:ACPower
Front Panel
Access: Shift-4 (Measure), ACPR, On/Off
Other Input Impedance Loss
[:SENSe]:CORRection:IMPedance[:INPut]:OFFSet
Description: Sets the value that is used for amplitude correction when the value set
by [:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude] is
something other than 50 or 75. This value is not applied if the
impedance is set to either 50 or 75.
Syntax: [:SENSe]:CORRection:IMPedance[:INPut]:OFFSet <rel ampl>
[:SENSe]:CORRection:IMPedance[:INPut]:OFFSet?
Cmd Parameter: <rel ampl>
Query Response: <rel ampl>
Range: 0 dB to 100 dB
Default Value: 0
Default Unit: dB
Related Command: [:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]
Front Panel
Access: Shift-8 (System), Application Options, Impedance