VNA Commands [:SENSe]:AVERage Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-103
3-21 [:SENSe]:AVERage Subsystem
This subsystem contains commands that are related to the combination of the data from
consecutive sweeps. Use commands in this subsystem to control sweep-to-sweep averaging
and max hold functionality.
Restart Averaging
[:SENSe]:AVERage:CLEar
Description: No query. Clears and restarts averaging of the measurement data. Note
that sweep averaging count must be set to greater than 1 for averaging
to restart.
Syntax: [:SENSe]:AVERage:CLEar
Cmd Parameter: <char>
Query Response: NA (no query)
Front Panel
Access: NA
Number of Traces to Average
[:SENSe]:AVERage:COUNt
Description: Sets the number of traces to be averaged. Note that when averaging
count is set to be greater than 1, sweep averaging is turned on. To stop
Syntax
[:SENSe]:AVERage:COUNt <integer>
[:SENSe]:AVERage:COUNt?
Cmd Parameter: <NR1> <integer>
Query Response: <NR1> <integer>
Range: 1 to 65535
Default Value: 1
Front Panel
Access: Shift-3 (Sweep), Sweep Averaging