VNA Commands :CALCulate<Tr>:FILTer[:GATE]:DISTance Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-11
Distance Domain Gate Display Settings
:CALCulate<Tr>:FILTer[:GATE]:DISTance:STATe
Description: Sets the gate display settings for the given trace. The gate can be either
OFF, in DISPlay mode, or ON. In DISPLay mode, the gate is shown on
the trace but is not applied to the transform. When set to ON, the gate
is shown and applied to the transform. <Tr> is the trace number in the
range of 1 to 4. If no trace number is specified, then the <Tr>
parameter defaults to trace number 1. Note that setting this also sets
the given trace as the active trace if it is not already active.
Syntax: :CALCulate<Tr>:FILTer[:GATE]:DISTance:STATe
OFF|DISPlay|ON
:CALCulate<Tr>:FILTer[:GATE]:DISTance:STATe?
Cmd Parameter: <char> OFF|DISPlay|ON
Query Response: <char> OFF|DISP|ON
Default Value: OFF
Example: To set Trace 2 to DISPlay view:
:CALC2:FILT:GATE:DIST:STAT DISP
Front Panel
Access: Freq/Time/Dist, Gate, Gate Function