VNA Commands :CALCulate:LIMit Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-29
Lower Limit Point Y Value
:CALCulate<Tr>:LIMit:LOWer:POINt:Y
Description: Sets the location of the lower limit point of the given trace <Tr> on the
y-axis at the specified location. <Tr> is the trace number in the range
1 to 4. If no trace number is specified then default is trace number 1.
Sending the set command changes the Move Limit on the front panel to
Point if it is currently set to Limit, and sets the given trace as the active
trace. The <y-parameter> is defined in the given trace current y-axis.
If no unit is specified with the <y-parameter>, then the default unit
is used. The query version of the command returns the location of the
given trace lower limit point on the y-axis. If an error occurs, such as
limit not ON, then the query version of the command returns
400 error codes. Limit line must be ON for the command to be valid.
Use :CALCulate:LIMit:TYPE to set the currently active limit line.
Syntax: :CALCulate<Tr>:LIMit:LOWer:POINt:Y <y-parameter>
:CALCulate<Tr>:LIMit:LOWer:POINt:Y?
Cmd Parameter: <NRf> <y-parameter> (depends on display type)
Query Response: <NR3> <y-parameter> (depends on display type)
Default Unit: Current active trace y-axis unit
Related Command: :CALCulate:LIMit:TYPE
[:SENSe]:TRACe<Tr>:SELect
:CALCulate<Tr>:FORMat
Front Panel
Access: Shift 6 (Limit), Limit Edit, Amplitude