Spectrum Analyzer Commands :CONFigure Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-29
6-4 :CONFigure Subsystem
This set of commands prepares the instrument for the selected measurement. It disables any
currently-enabled measurements and activates the specified measurement. It sets the
instrument to single sweep mode, waiting for an :INITiate command. It does not initiate
the taking of a measurement.
Current instrument settings may be changed to default values. These changes are identified
with their respective measurement commands.
Configure Adjacent Channel Power Ratio
:CONFigure:ACPower
Description: Configures the default adjacent channel power ratio measurement.
Disables any other active one-button measurements, including channel
power, occupied bandwidth, AM/FM demodulation and C/I. Sets the
main channel bandwidth equal to the span. Sets the adjacent channel
bandwidth and channel spacing equal to the main channel bandwidth.
Sets the detection method to RMS. Sets the instrument to single sweep
mode (:INITiate:CONTinuous OFF). Measurement settings can be
modified by using the [:SENSe]:ACPower commands before initiating
a sweep.
Syntax: :CONFigure:ACPower
Related Command: [:SENSe]:ACPower:STATe
[:SENSe]:ACPower:BANDwidth|BWIDth:MAIN
[:SENSe]:ACPower:BANDwidth|BWIDth:ADJacent
[:SENSe]:ACPower:BANDwidth|BWIDth:SPACing
Front Panel
Access: NA