VNA Commands :CALCulate:LIMit Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-47
Upper Limit X Value
:CALCulate<Tr>:LIMit:UPPer:X
Description: Moves the upper limit of the given trace <Tr> on the x-axis to the
given value. <Tr> is the trace number in the range 1 to 4. If no trace
number is specified, then the <Tr> parameter defaults to trace
number 1. <x-parameter> is defined in the given trace current
x-axis. The unit given with the <x-parameter> must correspond to
the given trace domain type. If no unit is specified with the
<x-parameter>, then the default unit is used. The set version of the
command changes the Move Limit on the front panel to Limit if it is
currently set to Point and sets the given trace as the active trace. The
query version of the command returns the location of the given trace
upper limit point on the x-axis followed by the unit. If an error occurs,
such as limit not ON, then the query version of the command returns an
error code of –400. Limit line must be on for the command to be valid.
Use the command :CALCulate:LIMit:TYPE to set the currently
active limit line.
Syntax: :CALCulate<Tr>:LIMit:UPPer:X <x-parameter>
:CALCulate<Tr>:LIMit:UPPer:X?
Cmd Parameter: <NRf> <x-parameter> (hertz, seconds, meters, feet)
Query Response: <NR3> <x-parameter> (hertz, nanoseconds, meters, feet)
Default Unit: Hz for Frequency domain,
Seconds for Time domain
Meters or Feet for distance domain.