Spectrum Analyzer Commands :CALCulate Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-5
Number of Limit Envelope Points
:CALCulate:LIMit:ENVelope:POINt
Description: Sets the number of inflection point for the currently active limit
envelope. Use :CALCulate:LIMit:TYPe to set the currently active
limit line.
Parameter: <number>
Syntax: :CALCulate:LIMit:ENVelope:POINt <number>
:CALCulate:LIMit:ENVelope:POINt?
Range: 2to41
Default Value: 21
Example: To set the number of inflection point to 31:
:CALCulate:LIMit:ENVelope:POINt 31
Front Panel
Access: Shift-6 (Limit), Limit Envelope, Upper Points (If Limit is toggled to
Upper)
Shift-6 (Limit), Limit Envelope, Lower Points (If Limit is toggled to
Lower)
Limit Envelope Shape
:CALCulate:LIMit:ENVelope:SHAPe
Description: Sets the currently active limit envelope shape.
Syntax: :CALCulate:LIMit:ENVelope:SHAPe SQUare|SLOPe
:CALCulate:LIMit:ENVelope:SHAPe?
Cmd Parameter: <char> SQUare|SLOPe
Query Response: <char> SQUare|SLOPe
Example: To set the limit envelope to a square:
:CALCulate:LIMit:ENVelope:SHAPe SQUare
Front Panel
Access: Shift-6 (Limit), Limit Envelope, Upper Shape (If Limit is toggled to
Upper) or Lower Shape (If Limit is toggled to Lower)