:CALCulate Subsystem Spectrum Analyzer Commands
6-22 PN: 10580-00306 Rev. E MS20xxC PM
Delta Marker X Value
:CALCulate:MARKer{1|2|3|4|5|6}:DELTa:X
Description: Sets the location of the delta marker on the x-axis at the specified offset
location, <x-parameter>, which is the offset value from the
reference marker position on the x-axis. <x-parameter> is defined in
the current x-axis units. The query version of the command returns the
location of the delta marker on the x-axis.
Syntax: :CALCulate:MARKer{1|2|3|4|5|6}:DELTa:X <x-parameter>
:CALCulate:MARKer{1|2|3|4|5|6}:DELTa:X?
Cmd Parameter: <x-parameter>
Query Response: <x-parameter>
Default Unit: Hz or seconds if in zero span
Example: If both the reference marker and delta marker #1 are currently at
2 GHz on the x-axis, send the command below to set the delta
marker #1 to 3 GHz on the x-axis (1 GHz offset from the reference
marker):
:CALCulate:MARKer1:DELTa:X 1GHz
In zero span, if both the reference marker and delta marker #1 are
currently at 35 microseconds on the x-axis, then send the following
command to set the delta marker to 60 µs on the x-axis (25 µs offset
from the reference marker):
:CALCulate:MARKer1:DELTa:X 25µs
Related Command: :CALCulate:MARKer[1]|2|3|4|5|6:X
Front Panel
Access: Marker, Delta
Delta Marker Read Y Value
:CALCulate:MARKer{1|2|3|4|5|6}:DELTa:Y?
Description: Reads the current Y value for the specified delta marker. The units are
the units of the y-axis.
Syntax: :CALCulate:MARKer{1|2|3|4|5|6}:DELTa:Y?
Default Unit: Current y-axis unit
Front Panel
Access: NA