VNA Commands :CALCulate:TRANsform Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-75
Time Domain Processing Mode
:CALCulate:TRANsform:TIME:TYPE:AUTO
Description: Sets the time domain transformation mode to either Auto (use
parameters ON or 1) or bandpass only (use parameters OFF or 0). In
Auto mode, the instrument uses lowpass time domain processing if the
instrument settings allow that (which typically occurs when the sweep
is a harmonic sweep). Otherwise, it uses bandpass processing. In
bandpass only mode (Auto OFF), the instrument forces the processing
to be always bandpass.
The query version of this command returns 1 if time domain processing
is currently set to automatic, or returns a 0 if Auto mode is set to OFF
(bandpass only mode).
Syntax: :CALCulate:TRANsform:TIME:TYPE:AUTO ON|OFF|1|0
:CALCulate:TRANsform:TIME:TYPE:AUTO?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: 1
Example: To set to Auto mode:
:CALCulate:TRANsform:TIME:TYPE:AUTO ON
Front Panel
Access: Shift-8 (System), Application Options, Time Domain, Domain Processing