[:SENSe]:CORRection:COLLect Subsystem VNA Commands
3-120 PN: 10580-00306 Rev. E MS20xxC PM
Calibration Steps and Calibration Types:
For each calibration Type, the following tables (Table 3-8, Table 3-9, and Table 3-10) list the
allowable calibration steps and port_no to be used in command
[:SENSe]:CORRection:COLLect[:ACQUire] <cal steps>,<port_no>. The
calibration steps are different for each calibration Method, and the port_no is different for
each calibration Type. For example, for calibration Type RFP1 and calibration Method SOLT,
the allowable <cal steps>,<port_no> settings are “OPEN,1”, “SHORT,1”, and
“LOAD,1”. The steps that are not allowed are indicated by “—”.
Table 3-8. SOLT Calibration Method
Calibration Type OPEN SHORT LOAD THRU ISOL
RF2P (Full 2 Port – S
11
, S
21
, S
12
, S
22
) 1,2 1,2 1,2 3 3
RFP1 (Full S
11
- Port1) 111— —
RFP2 (Full S
22
- Port2) 222— —
RFBP (Full S
11
& S
22
- Both Ports) 1,2 1,2 1,2 — —
TRFP (Response S
21
- Trans Response Fwd
Path)
— — —1 1
TRRP (Response S
12
- Trans Response Rev
Path)
— — —2 2
TRBP (Response S
21
& S
12
- Trans Resp Both
Paths)
— — —3 3
RRP1 (Response S
11
- Reflection Response
Port 1)
111— —
RRP2 (Response S
22
- Reflection Response
Port 2)
222— —
RRBP (Response S
11
& S
22
- Reflection
Response Both Ports)
1,2 1,2 1,2 — —
2PFP (1P2P S
11
, S
21
- 1 Path 2 Port
Fwd Path)
1 1 111
2PRP (1P2P S
22
, S
12
- 1Path 2Port
Rev Path)
2 2 222