EasyManua.ls Logo

Anritsu MS2026C - Field Strength Antenna; Field Strength Measurement State

Anritsu MS2026C
410 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
[:SENSe] Subsystem Spectrum Analyzer Commands
6-84 PN: 10580-00306 Rev. E MS20xxC PM
Field Strength Antenna
[:SENSe]:FSTRength:ANTenna
Description: Selects an antenna from the antenna list to use for field strength
measurement result calculations. The <antenna> argument is a
1-based index of the position of the desired antenna in the instrument
current antenna list. The list can be seen on the instrument by choosing
the Antenna soft key in the F Strength menu. For example, if the
desired antenna were the third item on the antenna listing, then the
value of the <antenna> argument would be 3. Setting the <antenna>
argument to 0 indicates that no antenna is selected (the query returns a
value of 0 (zero) for NO antenna).
The query form of this command returns the index of the
currently-selected antenna.
Syntax: [:SENSe]:FSTRength:ANTenna <antenna>
[:SENSe]:FSTRength:ANTenna?
Cmd Parameter: <antenna>
Query Response: <antenna>
Default Value: 1
Related Command: :CONFigure:FSTRength
Front Panel
Access: Shift-4 (Measure), Field Strength, Antenna
Field Strength Measurement State
[:SENSe]:FSTRength:STATe
Description: Sets the state of the field strength measurement, ON or OFF. When
using :CONFigure:FSTRength, the state is automatically set to ON.
Syntax: [:SENSe]:FSTRength:STATe OFF|ON|0|1
[:SENSe]:FSTRength:STATe?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF
Related Command: :CONFigure:CHPower
Front Panel
Access: Shift-4 (Measure), Field Strength, On/Off

Table of Contents