VNA Commands :MMEMory:LOAD Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-97
3-17 :MMEMory:LOAD Subsystem
The Mass MEMory LOAD subsystem contains commands to transfer from the mass memory
device to the internal memory.
Recall Setup
:MMEMory:LOAD:STATe
Description: No query. Recalls a previously stored setup from the current storage
location. The saved setup that is to be loaded is specified by
<file name>. <file name> must be enclosed in either single quotes
(‘ ’) or double quotes (“ ”) and must include the extension “.stp”.
The <integer> parameter is not currently used, but it must be sent.
Send a value of 1.
Syntax: :MMEMory:LOAD:STATe <integer>,<file name>
Cmd Parameter: <integer>,<string> (1, file name)
Cmd Parameter: NA (no query)
Related Command: :MMEMory:STORe:STATe
:MMEMory:MSIS INTernal|USB
Front Panel
Access: Shift-7 (File), Recall, Change Type (select file type from list)
Note
Recall and Save for both setup and measurement, as described in this section, are
specific for vector network analyzer modes, not for spectrum analyzer mode.
Note
When recalling a setup that causes a mode switch, wait a minimum of 60 seconds
before issuing the next command.

Table of Contents