VNA Commands :CALCulate:LIMit Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-45
Upper Limit Point Y Value
:CALCulate<Tr>:LIMit:UPPer:POINt:Y
Description: Sets the location of the upper limit point of the given trace <Tr> on the
y-axis at the specified location. <Tr> is the trace number in the range
1 to 4. If no trace number is specified then default trace is trace
number 1. Sending the set command changes the Move Limit on the
front panel to Point if it is currently set to Limit and sets the given trace
as the active trace. The <y-parameter> is defined in the given trace
current y-axis. If no unit is specified with the <y-parameter>, then
the default unit is used. The query version of the command returns the
location of the given trace upper limit point on the y-axis. If an error
occurs, such as limit not ON, then the query version of the command
returns an error code of 400. Limit line must be on for the command to
be valid. Use :CALCulate:LIMit:TYPe to set the currently active
limit line.
Syntax: :CALCulate<Tr>:LIMit:UPPer:POINt:Y <y-parameter>
:CALCulate<Tr>:LIMit:UPPer:POINt:Y?
Cmd Parameter: <NRf> <y-parameter> (depends on display type)
Query Response: <NR3> <y-parameter> (depends on display type)
Default Unit: Current active trace y-axis unit
Related Command: :CALCulate:LIMit:TYPe
:CALCulate<Tr>:LIMit:UPPer:POINt:X
[:SENSe]:TRACe<Tr>:SELect
:CALCulate<Tr>:FORMat
Front Panel
Access: Shift 6 (Limit), Limit Edit, Amplitude