[:SENSe]:CALibration Subsystem VNA Commands
3-104 PN: 10580-00306 Rev. E MS20xxC PM
3-22 [:SENSe]:CALibration Subsystem
This subsystem controls the system calibration.
Calibration State
[:SENSe]:CALibration:STATe?
Description: Reports the calibrated state. This command returns a 0 if there is no
valid calibration, otherwise it returns the bit of the S parameters that
has a valid calibration. The bits are as follows:
For example, if a value of 15 is returned, then all of the S parameters
bit is valid (because decimal 15 is equivalent to binary 1111). For
another example, if a value of 1 is returned, then S
11
has a valid
calibration because binary of 1 is 0x01.
Syntax: [:SENSe]:CALibration:STATe?
Cmd Parameter: NA (query only)
Query Response: <NR1> <integer> (0 to 15)
Front Panel
Access: NA
S11 bit 0x01
S12 bit 0x02
S21 bit 0x04
S22 bit 0x08