VNA Commands [:SENSe]:SWEep Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-143
3-30 [:SENSe]:SWEep Subsystem
This subsystem includes commands that affect the sweep parameters of the instrument.
IF Bandwidth
[:SENSe]:SWEep:IFBW
Description: Sets the IF Bandwidth. The <freq value> in Hz must be one of the
following 13 values:
100000|50000|20000|10000|5000|2000|1000|500|200|
100|50|20|10
The query form of this command returns the frequency in Hz.
Syntax: [:SENSe]:SWEep:IFBW <freq value>
[:SENSe]:SWEep:IFBW?
Cmd Parameter: <char> <freq value>
Query Response: <char> <freq value>
Range: 10 to 100000 Hz
Default Value: 10000
Default Unit: Hz
Example: Sets the IF Bandwidth frequency to 20 Hz:
:SENS:SWE:IFBW 20
Sets the IF Bandwidth frequency to 100 kHz:
:SENS:SWE:IFBW 100000
Front Panel
Access: Shift-3 (Sweep), IFBW