VNA Commands [:SENSe]:CORRection:COLLect Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-119
Calibration Steps
[:SENSe]:CORRection:COLLect[:ACQUire]
Description: Performs a measurement of the given steps. <cal steps> is the
calibration step to be performed and must be one of the following
values:
OPEN|SHORT|SHORT1|SHORT2|SHORT3|LOAD|
THRU|ISOLation
<port_no> is the port number, 1, 2, or 3. For calibration step OPEN,
SHORT, SHORT1, SHORT2, SHORT3, and LOAD, valid port number
is 1 or 2. For calibration step THRU and ISOLation, valid port
number is 1 for Fwd, 2 for Rev, and 3 for Fwd and Rev.
Note that the calibration step must be valid for the given calibration
type and calibration method. Refer to Table 3-8, Table 3-9, and
Table 3-10 for a list of valid calibration steps for each type and method.
The query version of this command returns a string that consists of the
last calibration step measurement that was performed followed by the
port number. The calibration step and port number are delimited by a
comma. Note that if no calibration step has been processed, then this
command returns the string “NONE, 0”.
Syntax: [:SENSe]:CORRection:COLLect[:ACQUire]
<cal steps>,<port_no>
[:SENSe]:CORRection:COLLect[:ACQUire]?
Cmd Parameter: <char> <cal steps>,<port_no>
Query Parameter: <char> <cal steps>,<port_no>
Query Response: <string>
Front Panel
Access: Shift-2 (Calibrate), Start Cal