VNA Commands :CALCulate:MARKer Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-55
Delta Marker State
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:DELTa[:STATe]
Description: Sets the specified marker as the active marker and turns it on or off. If
the value is set to ON or 1, then the specified marker is turned on and is
set as a delta marker. If the value is set to OFF or 0, then the specified
marker is turned off. The query version of the command returns a 1 if
the specified marker is a delta marker, and returns a 0 if it is not a delta
marker.
Syntax: :CALCulate:MARKer[1]|2|3|4|5|6|7|8:DELTa[:STATe]
OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:DELTa[:STATe]?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF
Example: To turn on marker #3 and set it as a delta marker:
:CALCulate:MARKer3:DELTa ON
:CALCulate:MARKer3:DELTa 1
:CALCulate:MARKer3:DELTa:STATe ON
:CALCulate:MARKer3:DELTa:STATe 1
To turn off delta marker #6:
:CALCulate:MARKer6:DELTa OFF
:CALCulate:MARKer6:DELTa:STATe OFF
:CALCulate:MARKer6:DELTa:STATe 0
Related Command: :CALCulate:MARKer:DELTa:REFerence