:CALCulate:LIMit Subsystem VNA Commands
3-52 PN: 10580-00306 Rev. E MS20xxC PM
Limit Y Value
:CALCulate:LIMit:Y
Description: Sets the location of the active limit line on the y-axis at the given value.
This moves the entire limit and moves the current active limit point by
the given value. Sending this command changes the Move Limit on the
front panel to Limit if it is currently set to Point. The <y-parameter>
is defined in the current y-axis. If no unit is specified with the
<y-parameter>, then the default unit is used. The query version of
the command returns the location of the active limit point on the y-axis.
If an error occurs, such as limit not ON, then the query version of the
command returns an error code of 400. Limit line must be on for the
command to be valid. Use :CALCulate:LIMit:TYPE to set the
currently active limit line.
Syntax: :CALCulate:LIMit:Y <y-parameter>
:CALCulate:LIMit:Y?
Cmd Parameter: <NRf> <y-parameter> (depends on display type)
Query Response: <NR3> <y-parameter> (depends on display type)
Default Unit: Current active trace y-axis unit
Related Command: :CALCulate:LIMit:TYPe
:CALCulate:LIMit:X
[:SENSe]:TRACe<Tr>:SELect
:CALCulate<Tr>:FORMat
Front Panel
Access: Shift 6 (Limit), Limit Edit, Amplitude

Table of Contents