VNA Commands [:SENSe]:CORRection:COLLect Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-123
Calibration Step Status
[:SENSe]:CORRection:COLLect:ACQUire:STATus?
Description: This command requests information about the current calibration step
or the specified calibration step. If no calibration step is specified, then
it returns a 1 if the current calibration step has completed, otherwise it
returns a 0. <cal steps> is the calibration step to be performed and
must be one of the following values:
OPEN|SHORT|SHORT1|SHORT2|SHORT3|LOAD|THRU|ISOLation
<port_no> is the port number, 1, 2, or 3. For step OPEN, SHORT,
SHORT1, SHORT2, SHORT3, and LOAD, valid port number is 1 or 2.
For calibration step THRU and ISOLation, valid port number is 1 for
Fwd, 2 for Rev, and 3 for Fwd and Rev.
Syntax: [:SENSe]:CORRection:COLLect:ACQUire:STATus?
[<cal steps>,<port_no>]
Cmd Parameter: NA (query only)
Query Parameter: <char> [<cal steps>,<port_no>]
Query Response: <NR1> <integer>
Front Panel
Access: NA