Spectrum Analyzer Commands :CALCulate Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-13
Set Limit Line Upper or Lower
:CALCulate:LIMit:TYPe
Description: Sets the currently active limit line to either upper or lower. Subsequent
limit line operations are performed on the selected limit line.
Syntax: :CALCulate:LIMit:TYPe 0|1
:CALCulate:LIMit:TYPe?
Cmd Parameter: <number>
0|1 (0 = upper limit line, 1 = lower limit line)
Query Response: <number>
0|1 (0 = upper limit line, 1 = lower limit line)
Range: 0|1
Default Value: 0 (upper)
Example: To set the active limit line to upper:
:CALCulate:LIMit:TYPe 0
Related Command: None
Front Panel
Access: Shift-6 (Limit), Limit
Upper Limit Alarm
:CALCulate:LIMit:UPPer:ALARm
Description: Enables/disables the alarm for the upper limit. It is a combination of the
commands :CALCulate:LIMit:TYPe 0;
:CALCulate:LIMit:ALARm ON|OFF. Setting the value to ON or 1
turns on the upper limit alarm. Setting the value to OFF or 0 turns off
the upper limit alarm. The query version of the command returns a 1 if
the upper limit line alarm is ON and returns 0 if OFF.
Syntax: :CALCulate:LIMit:UPPer:ALARm OFF|ON|0|1
:CALCulate:LIMit:UPPer:ALARm?
Cmd Parameter: <boolean> OFF|ON|0|1
Query Response: <bNR1> 0|1
Default Value: OFF
Related Command: :CALCulate:LIMit:ALARm
Front Panel
Access: Shift-6 (Limit), Limit Alarm