:MMEMory Subsystem AM/FM/PM Commands
7-10 PN: 10580-00306 Rev. E MS20xxC PM
7-6 :MMEMory Subsystem
The Mass MEMory subsystem contains functions that provide access to the instrument’s
setup and data storage.
:MMEMory:DELete <file name>
Title: Delete Setup/Measurement
Description: Removes the measurement or setup file specified by <file name> from
the current mass storage device. <file name> should be enclosed in
either single quotes (‘ ’) or double quotes (“ ”). It should contain one of
the following file extensions:
“.stp” for setup
“.spa” for SPA measurement
“.mwvna” for Vector Network Analyzer measurements
“.ia” for Interference Analysis measurements
“.cs” for Channel Scanner measurements
“.afp” for AMFMPM Analyzer measurements
Use the command MMEMory:MSIS to set the current mass storage
location.
Parameter: <file name>
Related Command: :MMEMory:STORE:STATe
:MMEMory:STORe:TRACe
:MMEMory:MSIS INTernal|USB
Front Panel
Access: Shift-7 (File), Delete, Delete Selected File
:MMEMory:LOAD:STATe <integer>,<file name>
Title: Recall Setup
Description: Recalls a previously stored instrument setup in the current storage
location. The setup file to be loaded is specified by <file name>. <file
name> should be enclosed in either single quotes (‘ ’) or double quotes (“
”) and should contain a file extension “.stp”. Use the command
MMEMory:MSIS to set the current storage location. The <integer>
parameter is not currently used, but it must be sent. Send a 1.
Parameter: <integer>, <file name>
Related Command: :MMEMory:STORe:STATe
:MMEMory:MSIS INTernal|USB
Front Panel
Access: Shift-7 (File), Recall