AM/FM/PM Commands :MMEMory Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 7-11
:MMEMory:LOAD:TRACe <integer>,<file name>
Title: Recall Measurement
Description: The instrument must be in the mode of the saved trace in order to recall
that trace.Use :INSTrument:SELect or :INSTrument:NSELect to set
the mode. Recalls a previously stored measurement trace from the
current storage location. The saved measurement trace to be loaded is
specified by <file name>. <file name> should be enclosed in either single
quotes (‘ ’) or double quotes (“ ”) and should contain a file extension.
Note that the trace specified by <file name> should be available at the
current mass storage location.
Use the command MMEMory:MSIS to set the current mass storage
location. The <integer> parameter is not currently used, but it must be
sent. Send a 1.
File name extensions:
“.spa” for SPA
“.mwvna” for Vector Network Analyzer
“.ia” for Interference Analysis
“.cs” for Channel Scanner
“.afp” for AMFMPM Analyzer measurements
Note: Extensions not available for T1 and Hi_PM.
Parameter: <integer>, <file name>
Example: To recall trace with file name “trace”:
:MMEMory:LOAD:TRACe 1,”trace.spa”
Related Command: :MMEMory:STORe:TRACe
:MMEMory:MSIS INTernal|USB
Front Panel
Access: Shift-7 (File), Recall Measurement
:MMEMory:STORe:STATe <integer>,<file name>
Title: Save Setup
Description: Stores the current setup into the file specified by <file name>. <file
name> should be enclosed in either single quotes (‘ ’) or double quotes (“
”) and should not contain a file extension. Use the command
MMEMory:MSIS to set the current storage location. The <integer>
parameter is not currently used, but it must be sent. Send a value of 0.
Parameter: <integer>, <file name>
Related Command: :MMEMory:LOAD:STATe
:MMEMory:MSIS INTernal|USB
Front Panel
Access: Shift-7 (File)