VNA Commands [:SENSe]:CORRection:CKIT:USER Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 3-117
DUT User Short Offset (SSST)
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SSST:SHORt[1]|2|3
Description:
Syntax: [:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SSST:SHORt[1]|2|3 <length>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SSST:SHORt[1]|2|3?
Cmd Parameter: <NRf> <length> (millimeters)
Query Response: <NR3> <length> (millimeters)
Example: To set the DUT short offset 2 for User 1 with line type coax and
calibration method SSST to 3 mm:
:SENSe:CORRection:CKIT:USER:COAX:SSST:SHORt2 3
Front Panel
Access: NA
DUT User Short Offset for calibration method SOLT
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide:SOLT
:SHORt
Description: Sets the DUT capacitance value for the specified user.
Syntax: [:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT:SHORt <length>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT:SHORt?
Cmd Parameter: <NRf> <length> (millimeters)
Query Response: <NR3> <length> (millimeters)
Example: To set the DUT short offset for User 1 with line type coax to 3 mm:
:SENSe:CORRection:CKIT:USER:COAX:SOLT:SHORt 3
Front Panel
Access: NA