Spectrum Analyzer Commands :READ Subsystem
MS20xxC PM PN: 10580-00306 Rev. E 6-51
Read Channel Power Density
:READ:CHPower:DENSity?
Description: Triggers a new channel power measurement and returns the channel
power density result. It is a combination of the commands :ABORT;
:INITiate; :FETCh:CHPower:DENSity? It returns only the channel
power density, not the channel power. Use the command
:READ:CHPower? to get both channel power and channel power
density. The channel power measurement must be the active
measurement (specified by :CONFigure:CHPower). The current
measurement can be queried using :CONFigure? command.
Syntax: :READ:CHPower:DENSity?
Cmd Parameter: NA (query only)
Default Unit: Current amplitude units
Related Command: :READ:CHPower?
:READ:CHPower:CHPower?
:CONFigure
Front Panel
Access: NA
Read Channel Power
:READ:CHPower?
Description: Triggers a new channel power measurement and returns the results. It
is a combination of the commands :ABORT; :INITiate;
:FETCh:CHPower? The channel power measurement must be active.
The current measurement can be queried using :CONFigure?
Syntax: :READ:CHPower?
Cmd Parameter: NA (query only)
Default Unit: dBm
Related Command: :READ:CHPower:CHPower?
:READ:CHPower:DENSity?
:CONFigure
Front Panel
Access: NA