:CONFigure Subsystem Spectrum Analyzer Commands
6-30 PN: 10580-00306 Rev. E MS20xxC PM
Configure Channel Power
:CONFigure:CHPower
Description: Configures the default channel power measurement. Disables any other
active one-button measurements, including ACPR, occupied bandwidth,
AM/FM demodulation, and C/I. Sets the integration bandwidth equal to
the span. Sets the detection method to RMS. Sets the instrument to
single sweep mode (:INITiate:CONTinuous OFF). Measurement
settings can be modified by using the [:SENSe]:CHPower commands
before initiating a sweep. Note that this measurement is not valid in
zero span.
Syntax: :CONFigure:CHPower
Related Command: [:SENSe]:CHPower:STATe
:SENSe:CHPower:BANDwidth|BWIDth:INTegration
Front Panel
Access: NA
Configure Field Strength
:CONFigure:FSTRength
Description: Configures the default field strength measurement. Disables any other
active one-button measurements, including channel power, adjacent
channel power, occupied bandwidth, AM/FM demodulation, and C/I.
Sets the antenna to the first antenna in the instrument’s antenna list.
Sets the instrument to single sweep mode (:INITiate:CONTinuous
OFF). Measurement settings can be modified by using the
[:SENSe]:FSTRength commands before initiating a sweep. Note that
this measurement is not valid in zero span.
Syntax: :CONFigure:FSTRength
Related Command: [:SENSe]:FSTRength:ANTenna